近赤外光による非侵襲的子宮内胎児機能診断システムの臨床評価

雅道 野川, M. Nogawa, 田中 志信 et al. · Kanazawa University Repository for Academic Resources (DSpace) (Kanazawa University) · 2015

Toward a development of non-invasive near-infrared assessment system for intrauterine fetal well-being, we have evaluated the performance of the proposed system, which offers non-invasive transabdominal measurement of fetal arterial oxygen saturation, with pregnant women in regular prenatal care. The proposed system consists of three major parts: a two-wavelength transabdominal fetal photo-plethysmography (750/830nm), frequency domain signal extraction procedure of fetal plethysmogram from transabdominal photo-plethysmogram, and non-blood drawing sensor calibration procedure based on radiative transfer for safe and reliable non-invasive fetal arterial oxygen saturation measurement. As a result of the clinical trials, the non-invasive fetal assessment system could offer acceptable normal arterial oxygen saturation ranges of intrauterine fetuses: 50-70%. 近赤外光による非侵襲的子宮内胎児機能診断システムの臨床評価にあたり,胎児血中酸素飽和度の非侵襲測定を実現するには,従来,事前に複数の胎児を対象に低酸素負荷時に採血試験を行い,光センサ出力との関係を校正する必要があった.しかしながら,このような校正法は危険を伴うため,光散乱理論を用い,妊婦腹部を模擬した数値モデルにおいて,子宮内胎児の光電容積信号(波長:750/830nm)から血中酸素飽和度を求める校正曲線を理論的(非採血的)に導出した.実際の妊婦腹部表面から得られた光電容積信号を用いて評価した結果,それぞれの酸素飽和度は生理学的な正常値(母体:98%前後,胎児:50~70%前後)を示した.

Cited by 0 works. doi:10.24517/00034484 Open-access PDF

Authors