トリチウムによる二次電子増倍管および質量分析計の機能障害とその回復
国昭 渡辺, 憲司 市村, Masao Matsuyama · Institutional Repositories DataBase (IRDB) · 1985
In D-T fusion experiments, secondary multipliers will be widely used as detectors for plasma diagnostics and mass spectrometers for fuel analysis and vacuum measurement. However, as significant impairment of the multiplier performance is predicted owing to the β-rays from tritium adsorbed on the detector surface (contamination), in-situ decontamination technique are, therefore, required. On reviewing the contamination effect on the performance of the secondary electron multipliers and decontamination techniques, it is indicated that the adsorption of tritiated water vapor is more serious than tritium gas. In principle, the tritium contamination could be removed by vacuum heating but it is a time consuming technique. Since photon irradiation is more effective in the decontamination than vacuum heating, it is proposed as an in-situ decontamination method. It is pointed out, however, that more extensive studies are required to understand the decontamination and/or signal discrimination techniques.
Cited by 0 works. doi:10.15099/00006335