ヘッドスペース法による水試料の放射性炭素測定のための前処理法の検討
治 高橋, 宙子 半田, Hiroshi Aoki Takahashi et al. · Institutional Repositories DataBase (IRDB) · 2014
We have established a sample preparation system for radiocarbon analysis of dissolved inorganic carbon (DIC) in water samples, using a headspace-extraction method. The system can be adapted for wide range of DIC concentrations. The water sample is injected to the vacuum vial by disposal syringe, and is acidified by H3PO4. Then, CO2 gas in the headspace flows out with He carrier gas to cryogenic trap. The purified CO2 yield is over 90% with He flow of 30-45 minutes. Extensive tests have shown that the extraction procedures do not introduce carbon isotopic fractionation and carbon contamination that can bias the 14C results.
Cited by 0 works. doi:10.18999/sumrua.25.180