推進剤流れ評価用希薄動圧計測器の小型化
澪 東, 峻佑 土金, 宜典 中山 et al. · Institutional Repositories DataBase (IRDB)
Since the electric propulsion propellant reflects on the wall surface in the vacuum chamber (ground test facility) and affects the thruster performance, some rarefied dynamic pressure measuring devices had been developed for the propellant flow estimation in our previous research.In this study, a restructured and miniaturized measuring device with a volume of approximately 27 cm 3 (a cube of 3 cm per side) was developed and fabricated to allow measurement in various narrow areas within the vacuum chamber.Through the preliminary test in the air and experiment within a vacuum chamber, it seems that the measurement resolution of the detected dynamic pressure is approximately 0.05 or 0.1 mPa and that the dynamic pressure distribution measured with this device was in general agreement with that measured with the previous device and that derived from a numerical simulation.