固相前処理/金薄膜電極アノーディックストリッピングボルタンメトリーによる土壌抽出液中ヒ素のオンサイト定量

義明 古庄, 伸明 牧田, 壮登 小野 et al. · Institutional Repositories DataBase (IRDB)

A method for the on-site estimation of arsenic (As) contamination in soil by solid-phase extraction with a resin based on molecular recognition technology (SPE-MRT) and anodic stripping voltammetry (ASV) using a gold film microelectrode was developed. Optimized pretreatments were carried out as follows: 1 g of soil sample was taken and put into polypropylene test tube containing 33 mL 1 M hydrochloric acid solution, which was heated at 70 degree Celsius for 5 min using graphite block hot plate. As in the sample was extracted by shaking for 1 min using an automatic mini-shaker system. The extract was finally filtered through a 0.45 μm disk PTFE membrane filter. The extracted arsenic was electrodeposited on the gold film carbon microelectrode at −400 mV vs. Ag/AgCl for 1 min and the deposits were stripped at a rate of 5 mV/s to 580 mV vs. Ag/AgCl by using AnaLig® TE-3, Cu in soil extract was removed with 1 M hydrochloric acid for the determination of As. The method can eliminate Cu interference. The detection limit was 0.5 μg/L in soil extract. The results of voltammograms for As in the extracts from real soil samples were the same level as these obtained by a cross-check method using inductively coupled plasma mass spectroscopy (ICP-MS).

Cited by 0 works. Open-access PDF

Authors