電磁流量計オンサイト校正法試験

等 荒木, 大山 信美, 荒 邦章 et al. · Institutional Repositories DataBase (IRDB) · 1984

In order to establish the cross correlation method as the in-situ calibration of the electromagnetic flowmeter, sodium flow rate measurement using the 6 inches permanent magnet type electromagnetic flowmeter was conducted. And it was found that the calibration was possible within the accuracy of $\pm$4\%. And useful method about the electrode positioning, signal processing and flow pattern correction were obtained for this purpose. This method will be applicable as the in-situ calibration method of "MONJU" main pipe flowmeter by the confirmation of this method on the larger pipe than this 6 inches' pipe.