透析器の物質除去性能に及ぼすモジュール形状の影響

梓 竹沢, 明泰 山下, Azusa Takezawa et al. · Institutional Repositories DataBase (IRDB) · 2009

Polysulfone (PS) is one of the most popular materials as a membrane for blood purification therapy. Solute removal performances of hemodialyzers, however, are dependent not only on physico-chemical structures of the membrane but also the design of modules or outer-casings. The authors evaluated four different PS membrane dialyzers in which two of them included the same membrane to demonstrate the importance of module housing. No significant differences were found in creatinine (MW 113) removal; however, higher solute removal was found in a dialyzer with a membrane of even lower solute permeability as molecular weight of a target solute became larger. Observed enhancement in solute removal may be caused by the internal filtration that was strongly dependent on the design of outer-casing. It was concluded that the module design of a dialyzer is as important as solute permeability of the membrane.

Cited by 0 works. Open-access PDF

Authors