新潟大学積雪地域災害研究センターにおける炭酸塩試料の炭素・酸素同位体比測定

隆 宮崎, 寿之 井川, 直喜 渡部 · Institutional Repositories DataBase (IRDB) · 2001

Measurements of carbon and oxygen isotope ratios for carbonate samples were started in earnest on a Micromass PRISM mass spectrometer in the Research Institute for Hazards in Snowy Areas, Niigata University. Carbonate samples were treated with 100% phosphoric acid at 190℃. The automatic front treatment device (MultiPrep) reduced the size of samples and cased the extraction and refining of CO_2 from carbonate samples. The δ^<13>C and δ^<18>O values of new laboratory standards are δ^<13>C_ =4.76‰ andδ^<18>O_ =-6.13‰ (FLD) and δ^<13>C_ =-32.19‰ and δ^<18>O_ =-22.44‰ (NCC), respectively. The new laboratory standards are effective in the correction of most unknown samples.

Cited by 0 works. Open-access PDF

Authors