光度滴定法による水溶液でのイオン性界面活性剤の自動定量

Kazuyuki DAITO, 達司 脇阪, Shinsuke Ishihara et al. · Institutional Repositories DataBase (IRDB) · 1997

A fully automatic photometric titration method was developed for the determination of anionic and cationic surfactants in a one-phase aqueous system with tetrabromophenolphthalein ethyl ester (TBPE) used as an indicator. To avoid decomposition of the TBPE indicator in the presence of an anionic surfactant in acidic solutions, the buffer and indicator solutions were automatically added to sample solutions just before titration. In an indirect titration method for cationic surfactants, cationic surfactants were automatically determined under the same conditions as in the titration of anionic surfactants. The usual samples were analyzed for less than 10 min at concentrations of 10(-6)10(-3) mol dm(-3). The relative standard deviations of the determination of various surfactants were 0.20.3%. The proposed method was applied to many practical samples. The results, except for those for anionic surfactants possessing polyoxyethylenyl groups and several cationic surfactants, were in good agreement with the results obtained by the JIS titration method (Epton method).

Cited by 0 works. Open-access PDF