心肺バイパスシステムにおける経時的静脈貯血量モニタリングの最適化:超音波・レーザー距離計測センサーシステムの開発と評価
秀暢 高橋, 寛 小林, 具挙 上月 et al. · Institutional Repositories DataBase (IRDB) · 2025
Currently, venous reservoir level sensors in cardiopulmonary bypass systems are limited to lower limit alarms to prevent air misfed and pump flow control, and are insufficient for safety improvement, automatic control, and appropriate management of blood volume. In this study, we developed a method to measure blood volume over time by measuring the distance to the liquid surface using small ultrasonic and infrared (Time of Flight) distance sensors in a venous blood reservoir. Using simulated blood, we proposed a modelshowing an exponential relationship between the distance to the liquid surface and the amount of blood stored and confirmed that both sensors were compatible with high accuracy (R² = 0.995 and 0.997, p < 0.01). The infrared method can achieve good measurement accuracy even in ambient light and bubbling, suggesting that it is effective in controlling blood volume over time.