軟部組織硬度計による生体硬度測定の信頼性の検討

Akira Takanashi, 亮 小沼, 琴美 塩田 et al. · Institutional Repositories DataBase (IRDB) · 2009

(PURPOSE)We assessed the soft tissue stiffness meter (STSM; TOKUSHUKEISOKU Corporation, TK-03C, Japan) for inter/intra-rater reliability. This device measures load and displacement simultaneously.(METHODS)1 healthy subject (with no history of orthopedic problem). Three physical therapists measured the surface stiffness of fore human body parts; intrinsic back muscles, upper trapezius, hamstrings and gastrocnemius of medial side. We each took 3 consecutive readings using the STMS at the right site on each of the human body parts. We calculated the displacement for a load of 5N and 10N. (RESULT)The overall rating reliability for all sites was high; coefficient of variation (CV)<5.0. The intra-rater reliability of the three therapists were high showing an intra-class correlation coefficient (ICC) (1・1) ≧0.95. The inter-rater reliability was also high; ICC( 2・1)= 0.98. (CONCLUSION)Thus, it seems that this STSM is a reliable and useful device for objective quantitative evaluation of soft tissue stiffness.