ディジタル乳房撮影における解像特性: System contrast transfer function の測定

KITAGAWA, Kaoru, Mizuki Hamano, Rie ISHII et al. · Institutional Repositories DataBase (IRDB) · 2021

System contrast transfer function (SCTF) has been used as an indicator of the image resolution properties in the quality assurance or quality control for digital mammography. We measured the SCTF using two test devices according to each procedure of test device, namely, the IEC and DMQC test devices. For unprocessed images, the SCTF values by the IEC test device were higher 4% at 2 cycles/mm and 6% at 4 cycles/mm than those by the DMQC test device. The causes are probably due to: the distance from chest wall side to bar pattern of the IEC test device is long, so the effective focal spot size for the IEC test device is smaller than that for the DMQC test device. In addition, as the compression paddle is removed in measurement using the IEC test device, scatter radiation from the paddle is little. The coeffi cient of variation (CV) of SCTF values by the IEC test device was smaller, 50% at 2 cycles/mm and 70% at 4 cycles/mm, compared to CV by the DMQC test device. Therefore, we concluded the IEC test device is more suitable than the DMQC test device for the SCTF measurement in the quality assurance or quality control because of its lower CV.