唾液アミラーゼモニターの唾液採取方法の違いによる アミラーゼ活性の比較 -唾液の回収が不十分と考えられる例と回収した唾液による活性測定の比較-

浩 棚橋 · Institutional Repositories DataBase (IRDB) · 2024

Nipro’s salivary amylase monitor is used in many clinical and educational settings, and allows the sensor chip to be placed under the tongue, saliva to be absorbed onto a collection paper, and the salivary amylase activity to be measured easily in about one minute. When activity was measured using this monitor in a basic biochemistry/biochemistry experiment with 160 students from 4 classes, 43 students (27%) had activity below 10 kIU/L. At the same time, when sublingual saliva was collected and placed directly on the chip to measure activity with the monitor, no samples showed activity below 10 kIU/L, and activity was 2.0 times higher than the method of collecting saliva by placing the sensor chip under the tongue. Furthermore, there was almost no correlation in activity between the two collection methods. When collecting saliva by placing the chip under the tongue, there is a possibility that not enough saliva is collected on the collection paper, and it may be necessary to measure activity using collected saliva when measuring in a group setting.