微小膜濃縮/黒鉛炉原子吸光法による鉛の高感度定量

光子 大島, 大崎, 隆史, 昌二 本水 · Institutional Repositories DataBase (IRDB) · 1998

Ultra-trace amounts of lead were determined by graphite-furnace AAS (GF-AAS) after miniature-membrane filter (MF) concentration/dissolution. Lead was concentrated as a pyrrolidinedithiocarbamate (APDC) chelate on a miniature MF, which was 5mm in diameter, 2mm in effective filtering diameter and 0.2 μm in pore size. After collection of the lead chelate on MF, the MF was cut so as to be 3mm in diameter using a punch. This tiny MF was dissolved in 100 μl of a dissolution solvent (methyl cellosolve), which resulted in a high concentration efficiency. When 10 ml of a sample was used, the concentration ratio was 100. The addition of 5 × 10(-5) M Zephiramine to the sample solution before chelation was effective for collecting lead; the recovery was almost 100%. The filtering time was 15 min for a 10 ml sample solution. The detection limit was 10 ppt when a 10 ml of sample solution was concentrated in 200 μl of methyl cellosolve. Measurements by GF-AAS were carried out using 40 μl of a methyl cellosolve solution. The relative standard deviation for 40 ppt of lead was 4.51% (n= 4). This method was applied to both river- and tap-water samples.