Yoshito Tsuji, 原, 俊明, 博 佐竹 et al. · Institutional Repositories DataBase (IRDB) · 1980

We tested a new time constant measuring method, field method, which can be applied to the T-sensor which is attached to any type of field instruments.This method can be executed in the field without special apparatus or testing tanks.The field method is effective wherever there is a thermocline with a large temperature gradient.In order to test the field method, the T-sensor time constant of the CTD was measured at the thermocline (0.14°C/m from 11 m to 20 m) in Tokyo Bay.The CTD is lowered across the thermocline rapidly, and raised slowly.The time constant can be calculated from the difference between the descending profile and the ascending profile, both of which are assumed to represent the identical temperature distribution.As the T-sensor speed increased, the time constant decreased until it settled at 0.4 sec.The time constant measured by the field method was larger than that measured by the step function response method.This difference may be attributed to the presence of the water boundary layer around the T-sensor probe in the field measurements.

Cited by 0 works. Open-access PDF