固相萃取-高效液相色谱荧光检测法测定人血浆3-硝基酪氨酸含量

郭帅, 邵丽华, 荣琳 et al. · 2009

A high performance liquid chromatography with fluorescence detection method has been developed for the determination of trace levels of 3-nitrotyrosine in human plasma. After precipitation of plasma proteins by acetonitrile and solid phase extraction, the derivatization reaction was carried out. The reaction mixture was determined by high performance liquid chromatography with fluorescence detection and quantified by external standard method. The coefficient correlation was 0.9999 in the linear range of 0.50-50.0 nmol/L. The average absolute and relative recoveries in spiked sample were 89.3%-91.9% and 99.2%-107.9%. The intra-day and inter-day relative standard deviations were 1.5%-4.8% and 2.8%-9.2%, respectively. The detection limit was 0.25 nmol/L. The average concentrations of 3-nitrotyrosine in normal human plasma (n=21) and patients with non-insulin dependent diabetes (n=23) were 1.18 and 3.63 nmol/L, respectively. The current method is sensitive, accurate, reproducible and suitable for analysis of large sample numbers in clinic research.