気管チューブ・カフに関する研究 その1 気管チューブ・カフの材質による差異の検討 : 笑気下における各種カフの内圧, カフ内ガス容積, ガス組成および濃度の変化について
幸雄 清水, 昭義 並木 · Institutional Repositories DataBase (IRDB) · 1981
The characteristics of endotracheal tube cuffs made of latex rubber (LR) or polyvinyl chloride (PVC) or silicone rubber (SR) were investigated in a model trachea and a vinyl bag filled with nitrous oxide and changes in intra- and lateral cuff pressure in a model trachea were measured. Increases in intracuff and lateral pressures in LR cuffs were more remarkable than those in PVC and SR cuffs. Changes in intracuff pressure, gas volume, component and concentration in air-inflated cuffs were measured in a vinyl bag filled with nitrous oxide. Nitrous oxide diffused into all cuffs in a concentration and time-related fashion. The rates of nitrous oxide diffusion in SR cuffs were very great compared with those in LR and PVC cuffs. SR cuffs were also more permeable to oxygen and nitrogen than LR and PVC cuffs. Gas volume in SR cuffs rapidly increased within a short time and then decreased, though intracuff pressures did not change from the control values. On the other hand, gas volume and intracuff pressure in LR and PVC cuffs increased continuously with time. These findings suggest that the overexpansion of SR cuffs and the high lateral pressure of LR cuffs may cause tracheal obstruction and histological damage to the tracheal wall.
Cited by 0 works. doi:10.15114/smj.50.351