Биомеханическая стандартизация чрескостного остеосинтеза
Барабаш Юрий Анатольевич, S. P. Shpinyak, Алфимов Роман Александрович et al. · Бюллетень Восточно-Сибирского научного центра Сибирского отделения Российской академии медицинских наук · 2011
The preference of pin-stem layout of external fixation device over the pin layout was proved. At the unstable fractures with load along the axis 5 times less durability was registered. In all cases regardless of fracture line (transverse, oblique, comminuted) we discovered increase of strength of fragments' fixation (10-85 % more) at the use of stem transosseous locks in intermediate bearings. Increase of fixation rigidity is observed at the increase in the direction perpendicular to the fracture. Choose of optimal direction of introduction of the shank in relation to the fracture line should conform to the supposed loading of patient and he character of fracture.